Isoperibol Calorimeter
Send a request
Isoperibol Calorimeter
The LEIC-6200 Isoperibol Calorimeter is a microprocessor-controlled oxygen bomb instrument for regular or occasional calorific testing. A general-purpose oxygen bomb and oval bucket fit inside the compact enclosure. This arrangement provides repeatable measurements. Both the bomb and bucket are removed and refilled by hand for each test, so operator involvement is greater than with an automatic fixed-bomb instrument.
The sensors, controls and jacket are integrated into a single cabinet. This self-contained assembly includes the following equipment and functions:
- A water jacket with temperature control, an internal circulation system and an electric heater.
- An oxygen bomb and oval bucket sized to fit within the insulating water jacket.
- An internal automatic oxygen-charging system for the bomb.
- An electronic thermometer for high precision temperature measurement.
- A bright color touch screen for entering data and controlling operation.
- Dedicated communication ports to connect a printer, computer and local area network (LAN).
- SD memory card slots for program updates and storage of archived test reports.
- A choice of combustion vessels for samples releasing from 52 calories to 12,000 calories in each test.
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Operating specifications
- Measurement method: isoperibol calorimetry.
- Both the oxygen combustion vessel and the bucket are removable.
- Test throughput: 4-7 tests per hour.
- Operator involvement is approximately 6 minutes per test.
- Instrument precision class: 0.05 to 0.1%.
- Temperature measurement resolution: 0.0001 ºC.
- Sample energy range: 52 to 12000 calorie, depending on the selected vessel.
- Linearity across the working range: 0.05%.
- SD memory storage and TCP/IP network communication.
- A USB Port supports connections to a balance and printer.
- Software updates are available through the Internet.
- Overall dimensions (in): 23w x 16d x 17h.
- Overall dimensions (cm): 57w x 40d x 43h.
Electronic temperature measurement
A high precision electronic thermometer measures the temperature using a dedicated thermistor sensor. The sensor is sealed within a stainless steel probe fixed in the calorimeter cover. Readings have a resolution of 0.0001° across a working range from 20° to 40°C. The instrument displays all temperatures in Celsius.
Temperature-controlled water jacket
An integral high capacity pump drives continuous forced circulation around the bucket chamber. Water flows along the chamber sides, beneath its bottom and through the cover. The jacket remains at a constant temperature for isoperibol operation, without requiring additional water or a waiting period at the end of a run. A sealed immersion heater and an internal heat exchanger regulate the jacket temperature. Both devices operate under the calorimeter controller.
Oxygen combustion vessel options
The supplied oxygen bomb accommodates samples releasing up to 8000 calories in each charge, using oxygen filling pressures up to 40 atm. For waste materials and chlorinated samples, an alternative bomb is recommended. This version provides greater resistance to chlorine and halogen acids generated during combustion.
Integral automatic oxygen charging
The calorimeter includes an automatic oxygen filling system to simplify and speed bomb preparation. Oxygen from a commercial cylinder connects to a solenoid inside the instrument, operated by a microprocessor. The operator attaches the filling hose connector to the bomb inlet valve and starts the sequence on the touch screen. The instrument then fills the bomb automatically, controlling the filling rate until it reaches the preset pressure.
Automatic energy equivalent calculation
The instrument calculates its energy equivalent (EE) from a series of standardization measurements. For each bomb and bucket combination, it averages an operator-selected number of the most recent standardization results. The default is 10 tests, as specified by ASTM and international methods. The operator can enter the largest acceptable standard deviation. If the results exceed that criterion, the calorimeter gives a notification. The computer can retain energy equivalent values for up to four different bomb and bucket combinations.
Adjustable operating options
Recommended procedures are provided, and the following settings allow the operator to adapt the measurement process:
- Program parameters can be changed for unusual sample sizes or particular precision requirements.
- The program accepts the titrant concentrations selected by the operator.
- Several measurement units are available for reporting calorific values.
- Acid correction handling can be adjusted to account for differences between ASTM, BS, DIN and ISO methods.
- Protection of program controls prevents unintended setting changes.
- A usage count can be kept for each bomb, with a notification when servicing is due.
Operating language selection
Programming choices, operating menus, reports and error messages can be displayed in English, French, German, Spanish, Italian, Polish, Chinese, Russian or Czech. The selected language applies to these interface and reporting functions.
Electronic transfer of data
A USB connection sends reports to a printer and receives sample mass readings from an analytical balance. Results are transferred to a laboratory computer through Ethernet using standard network communication protocols.
Modular construction and compatibility
Compatibility with oxygen combustion bombs and water handling equipment already in the laboratory is confirmed on request. Controller electronics are assembled on a removable chassis. This modular construction provides access for maintenance and service of the electronics.
Expanded calorimeter package
Items in the complete system package Component group Included equipment Measurement instrument Calorimeter Water circulation equipment Water handling system Report output device Printer with USB connection Additional combustion hardware Extra vessel Additional water container Extra bucket Vessel upkeep supplies Maintenance kit for the vessel Service supplies 1 Year service kit