Steady-State Gas Permeameter and Porosimeter
Send a request
CT-GP-PP
Steady-State Gas Permeameter and Porosimeter
The Steady-State Gas Permeameter and Porosimeter measures both gas permeability and porosity of plug-sized core samples in one bench instrument. It is a routine core analysis tool for service laboratories, operators and universities that need permeability by steady-state flow and porosity by isothermal nitrogen expansion on the same plug without moving it between instruments.
How it works
The plug is loaded in the core holder and confining pressure is applied to seal the sleeve against the sample. For permeability, gas is passed through the plug at a set rate on one of two flow ranges, 0 to 50 and 0 to 600 cc/min, and the differential pressure across the plug is recorded once flow is steady. Permeability is calculated from flow rate, differential pressure and plug dimensions. Flow pressure is 0 to 150 psi and the measured range covers 0.01 md to 10 darcies.
For porosity, a reference volume of nitrogen at known pressure is expanded into the pore space at constant temperature. The equilibrium pressure gives the pore volume, from which porosity is calculated. Pressure transducer accuracy is 0.1% of full scale and flow accuracy is 1% of full scale. The instrument operates at ambient temperature. Standard confining pressure is 0 to 400 psi; a 10,000 psi option is offered for measurement under overburden stress.
What is included
- Core holder for 1 inch diameter plugs up to 3 inches long (1.5 inch option)
- Confining pressure control, 0 to 400 psi standard
- Two gas flow ranges, 0 to 50 and 0 to 600 cc/min
- Pressure transducers with 0.1% F.S. accuracy
- Power supply for 110 to 220 VAC, 50 or 60 Hz
Specify for your project
- Plug diameter: 1 inch standard or 1.5 inch option
- Confining pressure: 400 psi standard or 10,000 psi option
- Permeability range of the samples within 0.01 md to 10 darcies
- Wetted materials, tooling and instrumentation, stated in the project datasheet
- Test gas available at the site, nitrogen or helium
- Line voltage and frequency
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Features
- Permeability: Range: 0.01md to 10 darcies
- Flow Pressure: 0 - 150 psi
- Confining Pressure: 0 - 400 psi (option 10,000 psi)
- Gas flow: 0-50 and 0-600 cc/min
- Temperature: Ambient Pressure transducer accuracy: 0.1% F.S.
- Flow accuracy: 1 % F.S.
- Core diameter: 1” and 1.5” (option)
- Core length: Up to 3”
- Power requirement: 110-220 VAC, 50 or 60 Hz