✓ In-house manufacturing — Hastelloy C276 & titanium✓ Custom engineering & prototypes✓ API / ASTM · CSA · CRN · CE✓ Serving US, Canada, EU, Middle East, Central Asia, Africa & LatAm

Mobile PVT System

  • Function. Performs PVT studies in the laboratory or on site under field conditions (dust, vibration, temperature variation) using an embedded high-pressure pump for pressure control.

    Controlled. Cell pressure (embedded high-pressure pump) and temperature.

    Measured. Phase volumes, saturation/bubble-point pressure.

    Configuration. Ruggedized for field deployment.

    Cell volume, pressure and temperature ratings, wetted materials, visual/measurement options and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.

    Model: LEPVTM

    • Mobile PVT System
  • Supported PVT procedures

    • CME: expansion at constant mass.
    • DV: differential vaporisation.
    • Separation by flashing.
    • CMD: depletion at constant mass.
    • CVD: depletion at constant volume.

    Cell operating data

    • The pressure rating is 15,000 psi.
    • Temperature operation starts at ambient and reaches 200°C (400 °F).
    • Temperature control maintains ± 0.5 °C.
    • The cell holds 300 cc, including a visible section of 20 cc.
    • Volume readings have an accuracy of 0.01 ml.
    • Pressure measurement accuracy is 0.1% of full scale.
    • Electrical input is 220 VAC at 50/60 Hz.

    Experiment functions

    • The design accommodates different PVT study requirements.
    • Operation uses no mercury.
    • The system identifies phases automatically and measures their volumes.
    • An integral pump regulates pressure precisely and measures displaced volume.
    • A video camera captures PVT tests as they take place.
    • Magnetic stirring helps the phases reach equilibrium quickly.
Coretest