Mobile PVT System
Send a request
CT-PVT-MOB
Mobile PVT System
Function. Performs PVT studies in the laboratory or on site under field conditions (dust, vibration, temperature variation) using an embedded high-pressure pump for pressure control.
Controlled. Cell pressure (embedded high-pressure pump) and temperature.
Measured. Phase volumes, saturation/bubble-point pressure.
Configuration. Ruggedized for field deployment.
Cell volume, pressure and temperature ratings, wetted materials, visual/measurement options and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.
Model: LEPVTM
Supported PVT procedures
- CME: expansion at constant mass.
- DV: differential vaporisation.
- Separation by flashing.
- CMD: depletion at constant mass.
- CVD: depletion at constant volume.
Cell operating data
- The pressure rating is 15,000 psi.
- Temperature operation starts at ambient and reaches 200°C (400 °F).
- Temperature control maintains ± 0.5 °C.
- The cell holds 300 cc, including a visible section of 20 cc.
- Volume readings have an accuracy of 0.01 ml.
- Pressure measurement accuracy is 0.1% of full scale.
- Electrical input is 220 VAC at 50/60 Hz.
Experiment functions
- The design accommodates different PVT study requirements.
- Operation uses no mercury.
- The system identifies phases automatically and measures their volumes.
- An integral pump regulates pressure precisely and measures displaced volume.
- A video camera captures PVT tests as they take place.
- Magnetic stirring helps the phases reach equilibrium quickly.