✓ In-house manufacturing — Hastelloy C276 & titanium✓ Custom engineering & prototypes✓ API / ASTM · CSA · CRN · CE✓ Serving US, Canada, EU, Middle East, Central Asia, Africa & LatAm

Video Tracker

  • Function. Tracks the phase interfaces of a three-phase mixture expelled from a core at reservoir-representative conditions using twin vertical bores.

    Controlled. Reservoir pressure and temperature.

    Measured. Phase-interface positions and phase volumes.

    Volume, pressure and temperature ratings, wetted materials, seals and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.

    Model: LEVT

    • Video Tracker
  • Technical characteristics

    • The tracker operates at ambient temperature.
    • Collection capacity is selectable as 100 ml or 300 ml.
    • Volume readings have an accuracy of 0.1 ml.
    • Glass forms the surfaces that contact fluid.
    • Provide 110-220 VAC electrical service at a frequency of 50 or 60 Hz.

    Operation and measurement

    • The imaging system detects multiple fluid interfaces.
    • Optical detection works independently of the type of fluid.
    • Volume calibration functions support accurate volume measurement.
Coretest