Two-Phase Video Separator (Ambient)
Send a request
CT-FM-VS2A
Two-Phase Video Separator (Ambient)
The Two-Phase Video Separator (Ambient), model CT-FM-VS2A, separates the liquid and gas phases produced during a core flood or other multiphase experiment and measures their volumes at atmospheric pressure. The separator body is glassware, and a color camera records the phase interface so that volumes are read visually with direct observation of the fluids. It belongs to the fluid monitoring group of special core analysis instruments, alongside the three-phase ambient unit, the HPHT video separators, the video tracker and the fraction collectors.
How it works
Effluent from the core holder enters the glass separator, where gas collects above the liquid. The camera has a 5 megapixel sensor and a 50 mm field of view, lit by a 20 W light source. Measurement resolution is 0.02 mm and volume accuracy is 0.01 mL. The standard maximum volume change is 50 mL; other volumes are available on request.
The unit operates at atmospheric pressure and up to 40 °C. The controlled condition is ambient separation. The measured variables are liquid and gas volume by visual reading. Power is 110 to 220 VAC, 50 or 60 Hz.
What is included
- Glass two-phase separator, 50 mL maximum volume change
- 5 megapixel color camera with 50 mm field of view
- 20 W lighting
- Other hardware and software: stated in the project datasheet
Specify for your project
- Volume change required, if different from 50 mL
- Fluids to be separated and the expected gas-to-liquid ratio
- Operating temperature, within the 40 °C limit
- Wetted materials and seals in contact with the effluent
- Interface to the flow system and to data acquisition
- Supply voltage and frequency
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Features
- Max pressure: Atmospheric pressure
- Max temperature: up to 40 °C
- Max volume change: 50 ml (other upon request)
- Volume accuracy: 0.01 ml
- Material: Glassware
- Measurement resolution: 0.02 mm
- Field of view: 50 mm
- CCD color camera: 5 megapixels
- Lighting: 20 watts
- Power: 110-220 VAC, 50 or 60 Hz