Gas Flow Control System
Send a request
CT-FM-GFC
Gas Flow Control System
The LEGFC gas flow control system injects pressurized nitrogen through a core sample at a constant flow rate. It connects to a core holder, controls gas pressure and flow, and measures the gas flow and pressure through the core. Core laboratories use it for gas permeability and gas injection tests at ambient temperature.
How it works
Nitrogen from a supply is regulated to the test pressure and passed through a flow controller set to a constant rate. The gas then enters the core holder inlet. Flow rate and pressure are the controlled variables; gas flow and pressure through the core are measured and recorded, so permeability can be calculated from steady-state flow and pressure drop. The system runs at ambient temperature and all wetted parts are stainless steel.
Versions and ranges
Version Maximum pressure GFC 3000 3,000 psi GFC 4350 4,350 psi Flow ranges are 0-10, 0-100 and 0-1,000 sccm; other ranges are available. A system can carry more than one range to cover tight and permeable samples.
Key data
- Gas: nitrogen
- Temperature: ambient
- Pressure: 3,000 psi or 4,350 psi
- Flow range: 0-10, 0-100 or 0-1,000 sccm, others available
- Wetted parts: stainless steel
Specify for your project
- Maximum injection pressure: 3,000 or 4,350 psi
- Flow range or ranges from expected sample permeability
- Core holder and connections the system feeds
- Data acquisition and software interface
- Nitrogen supply available at site
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Features
- Temperature: Ambient
- Pressure: GFC 3000: 3,000 psi GFC 4350: 4,350 psi
- Flow range: 0-10, 0-100 and 0-1,000 sccm, other ranges available
- Wetted parts: stainless steel