Hydraulic Intensifier
Send a request
CT-FS-INT
Hydraulic Intensifier
Function. Pressurizes fluid samples using a dual-chamber cylinder with a low-pressure and a high-pressure chamber separated by a dual-diameter piston.
Controlled. Low-pressure drive; output high pressure.
Measured. Output pressure and displaced volume.
Volume, pressure and temperature ratings, wetted materials, seals and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.
Models: LEHI, LEHI2, LEHI4, LEHI6, LEHI8, LEHI10, LEHI15
Intensifier configuration Pressure multiplication ratio High-pressure chamber volume (ml) High-pressure end rating (psi) High-pressure bore diameter (mm) High-pressure piston travel (mm) Low-pressure end rating (psi) Low-pressure bore diameter (mm) Ratio 2:1 configuration 2:1 50 ml 20,000 psi 25.46 mm 98.24 mm 10,000 psi 36 mm Ratio 4:1 configuration 4:1 25 ml 40,000 psi 18.00 mm 98.24 mm 10,000 psi 36 mm Ratio 6:1 configuration 6:1 20 ml 60,000 psi 14.70 mm 117.89 mm 10,000 psi 36 mm Ratio 8:1 configuration 8:1 15 ml 80,000 psi 12.73 mm 117.89 mm 10,000 psi 36 mm Ratio 10:1 configuration 10:1 10 ml 10,000 psi 11.38 mm 98.24 mm 10,000 psi 36 mm Ratio 15:1 configuration 15:1 5 ml 150,000 psi 9.30 mm 73.68 mm 10,000 psi 36 mm Installation and construction
- Use the intensifier at ambient temperature.
- The low-pressure port has a ¼ inch FNPT thread. The high-pressure port uses a ¼ inch FHP thread.
- Choose stainless steel or Hastelloy for the surfaces that contact fluid.
- Approximate mass is 22 kg; the assembly measures 600 mm in length.
- Body diameter varies with configuration, from 90 mm to 120 mm.