✓ In-house manufacturing — Hastelloy C276 & titanium✓ Custom engineering & prototypes✓ API / ASTM · CSA · CRN · CE✓ Serving US, Canada, EU, Middle East, Central Asia, Africa & LatAm

Hydraulic Intensifier

  • Function. Pressurizes fluid samples using a dual-chamber cylinder with a low-pressure and a high-pressure chamber separated by a dual-diameter piston.

    Controlled. Low-pressure drive; output high pressure.

    Measured. Output pressure and displaced volume.

    Volume, pressure and temperature ratings, wetted materials, seals and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.

    Models: LEHI, LEHI2, LEHI4, LEHI6, LEHI8, LEHI10, LEHI15

    • Hydraulic Intensifier
  • Intensifier configurationPressure multiplication ratioHigh-pressure chamber volume (ml)High-pressure end rating (psi)High-pressure bore diameter (mm)High-pressure piston travel (mm)Low-pressure end rating (psi)Low-pressure bore diameter (mm)
    Ratio 2:1 configuration2:150 ml20,000 psi25.46 mm98.24 mm10,000 psi36 mm
    Ratio 4:1 configuration4:125 ml40,000 psi18.00 mm98.24 mm10,000 psi36 mm
    Ratio 6:1 configuration6:120 ml60,000 psi14.70 mm117.89 mm10,000 psi36 mm
    Ratio 8:1 configuration8:115 ml80,000 psi12.73 mm117.89 mm10,000 psi36 mm
    Ratio 10:1 configuration10:110 ml10,000 psi11.38 mm98.24 mm10,000 psi36 mm
    Ratio 15:1 configuration15:15 ml150,000 psi9.30 mm73.68 mm10,000 psi36 mm

    Installation and construction

    • Use the intensifier at ambient temperature.
    • The low-pressure port has a ¼ inch FNPT thread. The high-pressure port uses a ¼ inch FHP thread.
    • Choose stainless steel or Hastelloy for the surfaces that contact fluid.
    • Approximate mass is 22 kg; the assembly measures 600 mm in length.
    • Body diameter varies with configuration, from 90 mm to 120 mm.
Coretest