✓ In-house manufacturing — Hastelloy C276 & titanium✓ Custom engineering & prototypes✓ API / ASTM · CSA · CRN · CE✓ Serving US, Canada, EU, Middle East, Central Asia, Africa & LatAm

Automated Confining Pressure and Temperature Controller

  • The Automated Confining Pressure and Temperature Controller (CT-CMP-ACPT, LEACP150 series) supplies confining fluid to a core holder at a set pressure and temperature and holds both at setpoint for the length of a test. It is one of the building blocks of coreflood and SCAL rigs and is suited to porous-media studies under NMR, CT scanning and in flow work.

    How it works

    A pump with a 250 cc cylinder pressurizes hydraulic oil to the confining pressure setpoint. The oil is circulated at up to 1 L/min and heated to the temperature setpoint, up to 150 °C, so confining pressure and sample temperature are controlled together from one unit. Confining pressure and temperature are measured and reported. Wetted parts are stainless steel, tubing connections are 1/4 in SLP, and power is 110 to 220 VAC, 50 or 60 Hz.

    Two models are listed; they differ only in pressure rating.

    ModelConfining pressurePore pressureTemperaturePump cylinderCirculation
    LEACP350-150up to 5,000 psiup to 5,000 psiup to 150 °C250 ccup to 1 L/min
    LEACP700-150up to 10,000 psiup to 10,000 psiup to 150 °C250 ccup to 1 L/min

    What is included

    • Pump with 250 cc cylinder for hydraulic oil
    • Confining pressure and temperature control with measurement of both
    • Circulation circuit, up to 1 L/min
    • 1/4 in SLP tubing connections

    Specify for your project

    • Pressure rating: 5,000 psi (LEACP350-150) or 10,000 psi (LEACP700-150)
    • Test temperature up to 150 °C
    • Core holder or holders to be served and number of samples
    • Imaging environment (NMR, CT scanning) or flow rig the controller connects to
    • Power supply, 110 to 220 VAC, 50 or 60 Hz
    • Seals, instrumentation and data interface per project

    This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.

    • Automated Confining Pressure and Temperature Controller
  • 5,000 psi temperature-controlled configuration

    • The controller reaches a temperature ceiling of 150°C.
    • Confinement is limited to 5,000 psi; the pore-pressure maximum is 5,000 psi.
    • Cylinder capacity for the pump is 250 cc.
    • Hydraulic oil circulates at rates no higher than 1 liter/min.
    • Connect tubing through the 1/4 inch SLP interface.
    • Supply voltage is 110-220 VAC. Either 50 Hz or 60 Hz is acceptable.
    • The parts exposed to operating fluid use stainless steel.

    10,000 psi temperature-controlled configuration

    • The controller reaches a temperature ceiling of 150°C.
    • Confinement is limited to 10,000 psi; the pore-pressure maximum is 10,000 psi.
    • Cylinder capacity for the pump is 250 cc.
    • Hydraulic oil circulates at rates no higher than 1 liter/min.
    • Connect tubing through the 1/4 inch SLP interface.
    • Supply voltage is 110-220 VAC. Either 50 Hz or 60 Hz is acceptable.
    • The parts exposed to operating fluid use stainless steel.