High-Pressure Flocculation Titrimeter
Send a request
CT-FA-HPFT
High-Pressure Flocculation Titrimeter
Function. Evaluates the precipitation and crystallization conditions of asphaltenes and paraffins in dark, undiluted crudes and oil products under pressure.
Controlled. Pressure, temperature and titrant addition.
Measured. Flocculation/precipitation onset.
Pressure and temperature ratings, wetted materials, cell/loop geometry and instrumentation are configured per project. Guaranteed performance and included hardware are stated only in the project-specific technical proposal and approved datasheet.
Model: LEFT
Technical data
- The testing method is ASTM D6703.
- The pressure limit is 10,000 psi; an optional version reaches 20,000 psi.
- Temperatures extend from -20°C to 200°C (400°F), and the vessel holds 100 ml.
- A magnetic mixer stirs the fluid. Fluid-contact parts use stainless steel, with Hastelloy offered as an option.
- Measurement uses light in the near-infrared band.
- The laser output is 250 mWatts.
- A sensitivity of 1 pWatt is specified for the detector.
- The detection dynamic range is 100 dB.
- Electrical input requires 220 VAC at 50/60 Hz on 1 phase.
Operation and applications
- Solid particles can be detected even in oils with very low API gravity.
- The complete instrument operates automatically.
- A stirrer located inside the fluid vessel uses a magnetic drive to mix the sample effectively.