Automated Asphaltene Extractor (ASTM D6560)
The Automated Asphaltene Extractor (CT-FA-EXT) automates the n-heptane method for asphaltene content in petroleum fluids according to ASTM D6560, IP 143 and NF 60115, reducing operator supervision compared with the manual procedure. Asphaltene content is reported as a mass fraction.
How it works
A sample of at least 1 g is treated with n-heptane, which precipitates the asphaltenes. The precipitate is separated by filtration on a medium with 0.8 µm particle retention and the asphaltenes are recovered with toluene; the unit runs the titration and separation steps automatically at temperatures up to 110 °C. The apparatus is glass and hermetically enclosed, so there are no toxic emissions and no loss of sample material, and the solvent is recycled, which lowers consumption. Solvent reservoirs hold 1 liter of toluene and 1 liter of n-heptane. Repeatability is 5% by weight or better, and the detection threshold is 0.3% by weight.
The extractor runs on 220 VAC, 50/60 Hz, single phase, and needs nitrogen at 30 psi. It is part of the asphaltene and solids detection line, alongside flocculation titrimeters, solid detection systems and high-pressure microscopes.
What is included
- Automated extractor in glass with hermetic enclosure and solvent recycling
Specify for your project
- Sample type and expected asphaltene content
- Number of samples per day
- Reference method to be reported: ASTM D6560, IP 143 or NF 60115
- Nitrogen and solvent supply at the installation site
- Supply voltage and frequency
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Technical data
- A sample mass of at least 1 gram is required.
- The specified toluene volume is 1 liter.
- The specified volume of N-heptane is 1 liter.
- The highest operating temperature is 110°C.
- The equipment uses glass as its construction material.
- Particle retention through the pores is specified at 0.8 µm.
- Electrical input uses 220 VAC at 50/60 Hz on 1 phase.
- The N2 supply must provide 30 psi.
Operation and applications
- Automatic measurement speeds up and simplifies the procedure specified by ASTM D6560-IP143.
- The stated repeatability is <= 5% Wt.
- Filtration supports a detection threshold of 0.3% Wt.
- A hermetically closed enclosure contains the sample and toxic vapours.
- Solvent recycling reduces the quantity consumed during testing.