Automatic Imbibiometer

  • The LEAIBBM is an automatic imbibiometer for rock wettability studies. It runs spontaneous imbibition and drainage tests on up to 12 core samples at once and records the volume of fluid each sample produces over time. Core analysis laboratories use it to build Amott-type production curves without manual burette readings.

    How it works

    Each core sits in its own cell: 6 imbibition cells and 6 drainage Amott cells. As the soaking fluid enters the pore space, the displaced fluid collects in a readable burette. A 5 megapixel color camera with 20 W lighting reads the burettes, and the system logs produced volume against time for every cell. Cells operate from ambient to 120 °C at up to 30 psi. Imbibition tests can also run in an accelerated mode.

    Specifications

    ParameterValue
    TemperatureAmbient to 120 °C
    PressureUp to 30 psi
    Sample diameterUp to 2 in
    Sample lengthUp to 4 in
    Imbibition cells6
    Drainage Amott cells6
    Burette readable volumeUp to 20 cc
    Fluid volume accuracy0.1 cc
    Camera5 megapixel color
    Lighting20 W
    Power supply110 to 220 VAC, 50/60 Hz

    Automated reading

    The camera replaces manual recording of burette levels. The operator no longer has to be present at each reading time, and the image-based reading improves the accuracy of produced volume against eye reading.

    Specify for your project

    • Number of samples per cell type, imbibition or drainage
    • Core diameter and length, up to 2 in by 4 in
    • Test temperature, up to 120 °C, and pressure, up to 30 psi
    • Test fluids, wetted materials and seals
    • Supply voltage

    This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.

    • Automatic Imbibiometer
  • Technical characteristics

    • The operating temperature extends from ambient conditions through 120°C.
    • The upper pressure limit is 30 psi.
    • Specimens can have diameters no greater than 2 inches.
    • The accepted specimen length reaches 4 inches.
    • The drainage arrangement contains 6 Amott cells.
    • A further 6 cells serve imbibition measurements.
    • The burette reading range reaches 20 cc.
    • Liquid volume readings have an accuracy of 0.1 cc.
    • A color CCD camera captures images at 5 Megapixels.
    • Illumination consumes 20 watts.
    • Provide 110-220 VAC electrical service at a frequency of 50/60 Hz.

    Operation and measurement

    • Controls support straightforward operation.
    • Camera-based readings improve measurement accuracy.
    • Automatic recording removes the need to transcribe readings by hand, reducing operator time.
    • Recorded imbibition tests can be retrieved in accelerated playback.