Aging Cell

  • Before special core analysis (SCAL), an aging cell conditions a core to help recover its original wettability. A laboratory sets the exposure period and maintains the required pressure and temperature while the sample ages. The standard arrangement contains one chamber for a single core.

    Sample conditioning

    A stainless steel chamber receives the specimen. Both pore pressure and confinement can reach 10,000 psi. Temperature control extends to 150 °C, while the permitted fluid throughput is 10 ml/min. The laboratory controls the aging time as well as the pressure and temperature. Its aging procedure determines how long these conditions are maintained before subsequent testing.

    Accepted specimen diameters are 1 and 1.5 inches. The available length accommodates cores through 6 inches. Operation requires nitrogen delivered at 2,000 psi, together with a 220 VAC electrical supply at 50 Hz.

    CharacteristicSpecified limit or requirement
    Pore pressure and confinement limit10,000 psi
    Temperature limit during aging150 °C
    Longest accepted specimen6 in
    Accepted specimen diameters1 in and 1.5 in
    Maximum fluid throughput10 ml/min
    Fluid-contact constructionStainless steel
    Nitrogen (N2) supply pressure2,000 psi
    Electrical service220 VAC, 50 Hz

    Standard arrangement

    • The standard assembly has one chamber for the core.
    • Fluid-contact components use stainless steel.
    • The confinement circuit and pore circuit each carry a 10,000 psi rating.
    • Aging temperature is controlled through 150 °C.

    Project inputs

    • Specify the required chamber count; the standard count is one.
    • Define the sample diameter as 1 or 1.5 in and its length within the 6 in limit.
    • State the required aging conditions, keeping pressure within 10,000 psi and temperature within 150 °C.
    • Identify the conditioning fluids and their required throughput, with a maximum of 10 ml/min.
    • Confirm the site electrical service and availability of nitrogen delivered at 2,000 psi.

    This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.

    • Aging Cell
  • Technical characteristics

    • Both pore pressure and confinement have a 10, 000 psi upper limit.
    • The cell controls aging temperature through 150°C.
    • Specimens may be as long as 6 inches.
    • Accepted diameters for the core are 1 inch and 1 ½ inches.
    • Fluid throughput has an upper limit of 10 ml/min.
    • Stainless steel forms the surfaces that contact fluid.
    • Provide nitrogen (N2) at a supply pressure of 2,000 psi.
    • Electrical service must deliver 220 VAC at a frequency of 50 Hz.