Lab Magnetic Stirring Synthesis Autoclave Reactor
Send a request
Lab Magnetic Stirring Synthesis Autoclave Reactor
The LELMSSAR is a small stainless steel synthesis autoclave with magnetic stirring, built for micro reaction tests in the laboratory. Volumes run from 25 ml to 1000 ml. Temperature and stirring speed are set and displayed digitally, and the vessel body separates completely from the heater for loading, cleaning and disassembly.
Design
The top cover is threaded onto the vessel. Inside, a PTFE magnetic stirrer turns at 60 to 2400 rpm. Operating pressure is below 6 MPa (60 bar), and the standard maximum temperature is 300°C, below 240 °C with a PTFE liner. The vessel is SS304 as standard; a PTFE liner, titanium or C-276 are optional.
Ports and fittings
- Gas port with ferrule-type needle valve for 6 mm pipeline
- Liquid port with needle valve and dip tube for sampling during the reaction
- Pressure gauge, 6 MPa range
- Rupture disc, 7 MPa opening pressure
- Temperature port with probe reaching deep into the vessel
- Exhaust valve for evacuating and purging gas before the reaction and venting after it
Models: LELMSSAR-25 (25 ml), LELMSSAR-50 (50 ml), LELMSSAR-100 (100 ml), LELMSSAR-150 (150 ml), LELMSSAR-200 (200 ml), LELMSSAR-250 (250 ml), LELMSSAR-300 (300 ml), LELMSSAR-500 (500 ml), LELMSSAR-1000 (1000 ml).
Specify for your project
- Volume, 25 to 1000 ml
- Maximum temperature and whether a PTFE liner is used
- Pressure, below 6 MPa
- Wetted material: SS304, titanium or C-276
- Power supply: 220V 50Hz or 110V 60Hz
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Structure characteristics:
- Stirring motor: High-power motor, powerful and stable.
- Inlet gas valve: gas inlet and outlet needle valve, connected to 6mm ferrule. The quality is reliable.
- Sampling valve: It is convenient to take samples at any time during the reaction and analyze the reaction process to connect 3mm or 6mm ferrules (optional).
- Exhaust valve: It is convenient to vacuum and replace the gas before the reaction. After the reaction, it is used as a vent valve.
- Stainless steel pressure gauge: real-time monitoring of reaction pressure.
- Temperature probe: deep inside the reactor body, real-time monitoring of reaction temperature.
Specifications:
Model LELMSSAR-25 LELMSSAR-50 LELMSSAR-100 LELMSSAR-150 LELMSSAR-200 LELMSSAR-250 LELMSSAR-300 LELMSSAR-500 LELMSSAR-1000 Volume 25ml 50ml 100ml 150ml 200ml 250ml 300ml 500ml 1000ml Max temperature Standard 300°C ( <240 °C For PTFE liner) Operate pressure <6Mpa (60bar) Material Standard SS304 (PTFE liner, titanium. C-276 optional) Stirring speed Inner PTFE magnetic stirrer; 60-2400rpm Top cover type Cover lid with thread connect Valve type Ferrule-type needle valve, connect with 6mm pipeline Opening port 1*gas port with needle valve; 1*liquid port with needle valve & dip tube; 1*Pressure gauge (6mpa range) : 1*rupture disc ( 7 mpa opening pressure); 1* temp. Measure port Voltage 220V 50Hz: 110V 60Hz optional