Touchscreen Reactor Controller for Stirred Reactors LERC-4848T
Send a request
Touchscreen Reactor Controller for Stirred Reactors LERC-4848T
The LERC-4848T Reactor Controller uses a touchscreen to operate stirred reactor control functions. It provides the capabilities of the standard modular Reactor Controller with its full complement of modules, presented through a touchscreen interface.
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
The fully populated controller includes primary temperature regulation, a motor control module, a pressure display module and a secondary temperature module. These functions are combined within the controller enclosure.
Primary Temperature Control Module (PTM)
A thermocouple or an RTD temperature sensor can provide the input to the primary module. Three outputs are assigned to heating control, cooling control and alarm-triggered disconnection of the heater. The temperature control function uses proportional, integral and derivative (PID) action and includes auto-tuning. Ramp and soak programming supports sequences of up to 64 steps.
Pressure Display Module (PDM)
The pressure module receives a signal from a transducer fitted on the reactor or on a connected accessory. It can be configured to cover a broad selection of operating ranges, and it displays the operating pressure continuously. Modules can be calibrated in psi, bar, or MPa. The pressure monitoring output operates an alarm relay. When pressure reaches the high limit selected by the operator, the relay switches off power to the heater.
Motor Control Module (MCM)
Closed loop feedback controls the stirring speed of the reactor. The primary module output continually adjusts motor voltage to respond to changes in motor load. This gives better regulation than the standard open loop speed control, especially when viscosity changes during a reaction. The output needed to hold a constant stirring speed also indicates how heavily the motor is loaded. As the load changes, the primary controller output changes with it. Although this value is not a direct torque reading, it can be used to follow reaction progress when the reaction produces a change in viscosity.
High Temperature Cut Off Module (HTM)
This limit controller adds a second temperature protection function to the main control module. Its redundant sensor may be installed inside the reactor or on its exterior. The primary output activates the lockout relay, shutting the system down for safety if the reactor reaches an unsafe temperature. The additional sensor supplements the main control measurement.
External Temperature Limit Module (ETLM)
In this configuration, the HTM sensor measures the reactor's outside wall temperature. Its primary output limits external heating: when the wall exceeds a predetermined temperature, the module interrupts the control signal from the main temperature controller. A secondary output activates the lockout relay in a non-latching mode if the wall reaches a preset unsafe temperature. This wall temperature limit provides an alternative to cascade control. It improves regulation in equipment with substantial thermal lag, including non-stirred reactors and reactors fitted with PTFE liners. It is also useful for low heat capacity reactants, including those used in gas phase reactions.
Solenoid Valve Module (SVM)
This package supplies a solenoid valve, a flow adjustment valve and all assembly parts for automatic coolant flow regulation through a reactor cooling coil. The valve package plugs into the controller cooling output socket. The intended coolant is tap water. Together, the included components form the automatic cooling flow control assembly.
Network connection
Network communications provide access to the controller from a tablet, a phone or a PC. After the controller has been connected to the network, these devices can communicate with it through a web server or a VNC server.
Controller enclosure dimensions Controller version Width, in. Height, in. Depth, in. Fully populated network controller 11.5 12.7 13.3