Reactor Controller for Non-Stirred Pressure Vessels
Send a request
Reactor Controller for Non-Stirred Pressure Vessels
The LERC-4838 Reactor Controller regulates temperature in non-stirred pressure vessels. It has no provision to control stirrer motor speed or operate cooling water, either in its current configuration or through later expansion.
The Primary Temperature Control Module is shared with the stirred-reactor controller and provides the same control and communication functions. A high/low heater switch is included, along with an overtemperature lockout relay and reset.
Control functions
- Programmable PID control with auto-tuning for accurate temperature regulation and reduced overshoot.
- Ramp and soak programs for temperature sequences.
- A heater switch selecting high or low power.
- A lockout relay with reset for overtemperature protection.
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Controller Modules and Expansion
The reactor controller offers a choice of four module types. Its enclosure accommodates the primary temperature control module and one additional expansion module. The extra position can take either a Pressure Display Module (PDM) or a High Temperature Cut Off Module (HTM). An External Temperature Limit Module (ETLM) is an alternative to the HTM. This arrangement allows the added function to be selected for pressure indication or temperature limiting.
This controller does not accept stirring control equipment. Modules associated with stirring are also outside its supported configuration.
Primary Temperature Control Module (PTM)
The primary module accepts temperature signals from either a thermocouple or an RTD sensor. It provides three outputs, allocated to heating, cooling and alarm-operated heater shutdown. Temperature regulation uses proportional, integral and derivative (PID) control. An auto-tune function is available to establish the control settings.Ramp and soak programs can contain up to 49 segments, allowing programmed changes of temperature and programmed holding periods within the controller sequence.
Pressure Display Module (PDM)
A pressure transducer on the reactor or on an attached accessory supplies the signal to this module. The input can be configured for a range of operating spans, and the module can be calibrated in psi or bar. Pressure readings are transmitted continuously to the connected PC. The pressure module output is connected to the alarm relay. If pressure reaches the high limit entered by the operator, that relay removes power from the heater.High Temperature Cut Off Module (HTM)
This separate limit controller supplements the primary temperature controller. It uses a redundant temperature sensor, which can be positioned inside or outside the reactor. The main module output activates the lockout relay. If the reactor reaches the unsafe temperature limit, the relay provides a protective shutdown.External Temperature Limit Module (ETLM)
The ETLM uses the same module hardware as the HTM, with the sensor positioned to measure the outer reactor wall. Its primary output restricts external vessel temperature. When the outside temperature passes a predetermined value, the module interrupts the control signal from the primary temperature controller.A secondary ETLM output activates the lockout relay without latching if the wall reaches a preset unsafe temperature. This arrangement offers an alternative to cascade control. It can improve temperature regulation where a large thermal lag separates heat input from the process response, including non-stirred reactors and vessels with PTFE liners. It also applies where reactants have low heat capacity, as in gas-phase reactions.
PC Communication and Data Collection
Each controller includes software for communication in both directions between the instrument and a connected PC. The PC can send operating set points and control parameters to every installed operating module. In the other direction, the software transfers the data collected by those modules to the PC for display and logging. The communication arrangement therefore covers both module configuration and collection of operating measurements.
The included logging software requires an RS-485 to USB conversion cable for operation. That cable provides the connection used by the PC communication arrangement.
For straightforward data logging, the included software provides the required functions when used with the RS-485 to USB conversion cable. An optional human-machine interface software package is recommended when the operator needs to change settings from the interface and record several variables. This package extends the user interface beyond basic collection of process measurements.
Controller Dimensions Controller Type Width, in. Height, in. Depth, in. Reactor controller 10.19 7.11 7.92 Optional Configurable Operator Interface
The optional PC software provides an adaptable user interface for process controllers. A common Modbus communication architecture also permits its use with the compatible reactor controller families. This shared communication method supports interface customization and access to the operating information from the connected instruments.
The controller-compatible software version supports up to eight individual control loops or limit controllers. Supported controller combinations are provided on request. Compared with the standard logging software, the optional interface provides the following functions:
- More detailed real-time plotting, with multiple charts and no limit on the number of plotted pens within each chart.
- Logging of any controller parameter, extending data collection beyond the process variable alone.
- Recording of alarms and events in the log.
- More prominent indication and annunciation of alarms.
- An interface that can be customized to the requirements of the particular application.
- Localization of the operator interface, including support for different languages.
Supported Operating Systems
The operator interface software supports the following Windows versions and system architectures:
- Windows 10, 8, 8.1, & 7; 32 bit & 64 bit.
Windows 8 RT is not supported. This restriction also applies to tablets using that operating system; the operating system version determines compatibility. Compatibility with the software version supplied is confirmed on request.
Minimum Computer Configuration
- A processor that meets the requirements of the Windows version in use.
- 4GB RAM minimum. 8GB or more is recommended because Windows consumes much of the minimum memory allocation.
- 1GB of hard disk capacity for installation of the operator interface and its associated utilities.
- Additional disk capacity for logged data. The required capacity depends on the amount of data collected and the logging frequency.
- A display adapter supporting 1024 x 768 pixels with at least 16-bit color.
- 2 USB ports for the required connections.
Preferred Computer Configuration
- A processor running at 2.6 GHz.
- 8GB RAM for the preferred configuration.
- A 500 GB hard disk, with sufficient further capacity for the required data logging.