Dean-Stark Extractor

  • The LEDSE Dean-Stark extractor determines fluid saturation in a core sample by solvent distillation and extraction. Routine core analysis laboratories use it to measure water saturation and residual fluid saturation on plug and full-size core. The core sits in the extractor body above a boiling flask; solvent vapor passes through the sample, condenses and refluxes back, and water removed from the pore space collects in a graduated receiving tube.

    How it works

    The boiling flask heater reaches up to 450°C. A water-cooled condenser returns the solvent; the cooling water supply is 0.5 to 3 liters per minute at 18°C. The controlled parameters are solvent reflux and heating. The water volume read on the receiving tube gives the water saturation. Power supply is 220 VAC 50 Hz for both versions.

    Two versions

    ParameterPlug coreFull-size core
    Core diameterup to 1.5"up to 4"
    Core lengthup to 4"up to 6"
    Extractor capacity250 ml3 liters
    Boiling flask500 ml6 liters
    Receiving tube5 and 15 ml, 0.1 ml graduation40 ml, 0.2 ml graduation, drain valve

    The plug version uses small receiving tubes with fine graduation for the low water volumes held in a plug. The full-size version handles whole core sections and adds a drain valve on the receiving tube, so collected water can be removed without dismantling the glassware.

    Specify for your project

    • Core size: plug up to 1.5" x 4" or full-size up to 4" x 6"
    • Receiving tube volume and graduation
    • Number of extraction positions required
    • Cooling water supply, 0.5 to 3 liters per minute
    • Power supply, 220 VAC 50 Hz as listed

    This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.

    • Dean-Stark Extractor
  • Dean-Stark extractor for plug core samples:

    • Electrical supply: 220 VAC, 50 Hz
    • Water cooling: 0.5 to 3 liters per minute at 18°C
    • Heating temperature: up to 450°C
    • Receiving tubes: 5 ml and 15 ml, graduated in 0.1 ml
    • Boiling flask: 500 ml
    • Extractor capacity: 250 ml
    • Core sample length: up to 4 in
    • Core sample diameter: up to 1.5 in

    Dean-Stark extractor for full-size core samples:

    • Electrical supply: 220 VAC, 50 Hz
    • Water cooling: 0.5 to 3 liters per minute at 18°C
    • Heating temperature: up to 450°C
    • Receiving tube: 40 ml, graduated in 0.2 ml, with drain valve
    • Boiling flask: 6 liters
    • Extractor capacity: 3 liters
    • Core sample length: up to 6 in
    • Core sample diameter: up to 4 in
Coretest