Digital Density Meter
Send a request
CT-DEN-DIG
Digital Density Meter
LEDDM provides digital density readings for fluids under reservoir-analysis pressure and temperature conditions. Its measuring chamber uses Hastelloy to handle high pressures and temperatures. In a PVT laboratory, the meter measures live oil, brine, gas condensates and other reservoir fluids across the conditions used for fluid analysis.
Measurement procedure
Sample enters the chamber through fittings sized 1/8 inch. The system establishes the required pressure and temperature, holds those settings and measures density under those conditions. Keeping the sample at line conditions prevents a flash to atmospheric pressure. The resulting reading therefore represents the fluid in its reservoir state or under the conditions maintained inside a PVT cell.
Measurement specifications
Measured or specified item Operating value Temperature limit up to 200 °C Pressure limit up to 20,000 psi Measurable density 0 to 3 g/cm3 Density measurement accuracy ±1 x 10-3 g/cm3 Connection size 1/8 inch Fluid-contact metal Hastelloy Electrical input 220 VAC, 50 Hz Laboratory applications
- Measure the density of live fluids during PVT experiments.
- Determine densities of pressurized brines and injection fluids.
- Measure density changes with pressure and temperature to support equation-of-state tuning.
Project information required
- State the highest required pressure and temperature, within the limits of 20,000 psi and 200 °C.
- Identify the fluids and their H2S or CO2 content so that material compatibility can be checked.
- Specify how the instrument will connect to a sample cylinder or PVT cell.
- Define the data interface required by the PVT software.
- State any electrical supply requirement that differs from 220 VAC, 50 Hz.
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Density measurement specifications
- The cell operates at temperatures reaching 200°C.
- The pressure capability extends up to 20,000 psi.
- Measurable densities span 0 to 3 g/cm3.
- The stated density accuracy is 10-3 g/cm3, also expressed as +/- 1 x 10 -3 g/cm3.
- Connections use fittings sized 1/8 inch.
- Hastelloy forms the surfaces exposed to the fluid.
- Electrical input is 220 VAC at 50 Hz.