Dynamic scale loop
Send a request
LE/SCALEVAL
Dynamic scale loop
The dynamic scale loop examines mineral precipitation and deposit formation at the pressures and temperatures encountered in reservoirs and pipelines. Tests assess how chemical inhibitors prevent mineral scale, including calcium carbonate, calcium barium and barium sulfates. The equipment determines the minimum inhibitor concentration (MIC) needed to stop deposits and compares different inhibitors under matching experimental conditions.
The method uses tube blockage, an established industry test, to determine MIC while fluid flows dynamically at a fixed temperature. Pressure loss along the pipe indicates the deposition rate during scale and pre-scale tests and provides the basis for assessing inhibitor performance. Inhibitor efficiency is calculated by dividing the time to blockage with inhibitor present by the ordinary blockage time. The operator can run experiments manually or select automatic operation.
This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.
Technical data
- The maximum pressure is 6,000 psi.
- The flowing fluid can reach a rate of 10 ml/min.
- Temperature measurement accuracy is 0.1°C.
- The materials used are Hastelloy, Monel and PTFE.
- Electrical input may be 110 or 220 VAC at a frequency of 50/60 Hz.
Operation and applications
- The full test sequence can run automatically.
- The system measures the performance of inhibitors against scale.
- Flowing tests reproduce conditions encountered in pipelines.