✓ In-house manufacturing — Hastelloy C276 & titanium✓ Custom engineering & prototypes✓ API / ASTM · CSA · CRN · CE✓ Serving US, Canada, EU, Middle East, Central Asia, Africa & LatAm

HPHT Interfacial Tension Meter

  • The HPHT Interfacial Tension Meter (model CT-EOR-IFT) determines the interfacial tension between two fluids, or the surface tension of a single fluid, at reservoir-representative pressure and temperature. It is used in EOR and reservoir fluid studies where fluid-fluid tension has to be known at the conditions of the reservoir rather than at ambient conditions.

    How it works

    The test fluids are loaded into a measuring cell rated to 700 bar (10,000 psi). Cell pressure and cell temperature are the controlled variables. The temperature range runs from ambient to 180 °C with a stated accuracy of 0.1 °C. A drop of one fluid is formed in the second fluid inside the cell, and the drop image is processed by the drop analysis software, which returns the interfacial or surface tension. The standard measuring range is 0.1 to 72 mN/m.

    Wetted parts are stainless steel as standard. A Hastelloy option is listed. The instrument is supplied for a 220 VAC, 50 Hz supply.

    What is included

    • Pressure and temperature controlled measuring cell, 700 bar, ambient to 180 °C
    • Drop analysis software
    • Stainless steel wetted parts (Hastelloy as an option)

    Specify for your project

    • Fluid pairs to be tested, to select stainless steel or Hastelloy wetted parts
    • Required pressure and temperature envelope within the 700 bar and 180 °C ratings
    • Expected interfacial tension range within 0.1 to 72 mN/m
    • Electrical supply at site if different from 220 VAC, 50 Hz
    • Additional modules or instrumentation required for the study

    This description defines the product family. Guaranteed performance, included hardware, exclusions, interfaces and acceptance criteria are stated only in the project-specific technical proposal and approved datasheet.

    • HPHT Interfacial Tension Meter
  • Features

    • IFT standard measurement : 0.1 to 72 mN/m
    • Temperature : Ambient to 180°C
    • Temperature accuracy : 0.1 °C
    • Pressure : 700 bar (10,000psi)
    • Wetted parts : Stainless steel, option: hastelloy
    • Power supply: 220 VAC 50 Hz

    Benefits

    • Easy to use
    • Provided with an advanced drop analysis software
    • Excellent drop stability