Molecular weight tester LEMWT

  • This instrument yields the liquid sample’s total molecular weight allowing the fraction of components heavier than C36+ undetected by the HC analyzer to be deduced. This laboratory model automatic cryoscope allows molecular weight, solution concentration, and freezing point depression determinations for 2.0 ml aqueous and various solvent solutions.

  • Features

    • Cell Temperature Range : -6 to +26°C
    • Operation : automatic or manual
    • Speed : 2 minutes per measurement
    • Precision temperature : 0.001°C, 1S.D. depends of solvent used
    • Precision molecular weight : 1% typical, depends of solvent used
    • Calibration : Two non interacting controls
    • Bath : 55 ml 15 minute cool dowm from ambient temperature
    • Power : 110-220 VAC, 50 or 60Hz
    • Weight : 13.6 kg
    • Dimensions (mm) : 35.5 x 20.3 x 25.4 cm
    • Molecular weight tester LEMWT