Summary. The Vinci RC4500 and the Coretest LERC4500 are the same class of instrument: a refrigerated, low-speed, long-radius capillary pressure centrifuge with six core holders, a 4,500 rpm ceiling, 5 to 90 °C and a camera that reads every receiving tube during rotation. Both offer overburden core holders to 5,000 psi. The RC4500 adds a published pore-pressure module to 5,000 psi and a 208 V / 60 Hz supply option. The LERC4500 has a longer rotor arm (26 cm against 23 cm), plugs up to 4 in long in atmospheric holders and a published camera resolution.

Why laboratories compare

Laboratories look for an alternative to the RC4500 when they need a second qualified source for a six-place refrigerated centrifuge, when delivery time or budget rules out the incumbent, or when the test programme calls for longer plugs, larger receiving tubes or a specific holder arrangement written into the project specification.

Coretest Capillary Pressure Refrigerated Centrifuge LERC4500 vs Vinci RC4500

Coretest Capillary Pressure Refrigerated Centrifuge LERC4500Vinci RC4500
Core holders6Up to 6
Plug diameter1 in, 30 mm or 1.5 in1 in, 30 mm or 1.5 in
Plug lengthUp to 4 in in atmospheric holders; up to 2.5 in in overburden holdersNot stated
Receiving tubes10 mL and 32 mL, calibrated13 cc and 28 cc
Speed range50 to 4,500 rpm, ±1 rpm, speed ramp50 to 4,500 rpm, ±1 rpm, step overshoot ≤5 rpm, acceleration 1 to 1,000 rpm/min
Rotor radius / RCFOuter radius 26 cm; up to 5,818 ×gLarge radius 23 cm; RCF not stated (about 5,200 ×g calculated at 4,500 rpm)
Temperature5 °C (at standstill) to 90 °C, two at-sample sensors5 °C (at 0 rpm) to 90 °C
ReadoutDigital camera, 5488 × 3672 px; interface better than 0.01 mL on a 10 mL tubeDigital camera synchronized with the tachometer, all six holders imaged at every rotation; resolution not stated
OverburdenOption: hydrostatic holders, 5,000 psi confining, 1,450 psi pore pressure, set before the run; no release between drainage and imbibitionOption: holders to 5,000 psi; switch from drainage to imbibition without releasing confining pressure
Pore pressure modulePore pressure up to 1,450 psi in the overburden optionSeparate module: overburden and pore pressure up to 5,000 psi each
Interpretation softwareAcquisition software converts interface position to produced volume; production versus time for relative permeabilityHassler–Brunner, Forbes, Hagoort; curve fitting; numerical history matching for Pc and Kr
SafetyImbalance, rotor identification, overspeed, over-temperature, lid interlockRotor imbalance sensor
Power400 V, 3 phase + N400 V 3~ + N (50/60 Hz) or 208 V 3~ + N (60 Hz)
Size / weight973 × 813 × 1015 mm / 450 kgNot stated
SupplyConfigured to orderCatalog product

“Not stated” means the figure is not given in the manufacturer's published material; it does not mean the feature is absent.

What the differences mean in practice

Capillary pressure range. At the same 4,500 rpm the longer 26 cm arm of the LERC4500 gives about 13 % more centrifugal acceleration at the outer face than a 23 cm arm, and for a 2 in plug about 15 % more inlet-face capillary pressure at the same speed. Both machines stay in the low-speed class, where speed holds steadily and the camera reads a clean interface.

Plugs and tubes. Both accept 1 in, 30 mm and 1.5 in plugs. The LERC4500 publishes plug length up to 4 in and a 32 mL tube, useful for long, high-porosity plugs where produced volume exceeds a 28 cc tube.

Reservoir conditions. Both offer 5,000 psi overburden holders. If your study needs pore pressure above 1,450 psi, for example CO2 or H2 storage work at reservoir pore pressure, the RC4500 pore-pressure module publishes 5,000 psi and is the closer fit today.

Software. Vinci publishes a full interpretation package with Hagoort and history matching. The LERC4500 delivers the production data per holder and step; many laboratories already run their own Pc and Kr interpretation and only need clean data.

Who each is for

Coretest LERC4500
A six-place refrigerated capillary pressure centrifuge configured to your plug sizes and tubes, with a long rotor arm, a high-resolution camera and optional overburden holders; or a second source next to an existing RC4500.
Vinci RC4500
Laboratories that need pore pressure to 5,000 psi at reservoir temperature, a 208 V / 60 Hz supply, or an integrated interpretation package from one vendor.

Moving an existing RC4500 method

Send your speed schedule, plug dimensions, fluid pair and tube sizes. Coretest recalculates the step speeds for the 26 cm rotor so each step gives the same inlet-face capillary pressure as your current method, and flags any step where tube volume or plug length needs a different holder.

Questions

Can the LERC4500 reproduce RC4500 capillary pressure steps?

Yes, by speed. Pc depends on rotor radius and speed; with a 26 cm arm the same Pc is reached at a slightly lower rpm. The step table is recalculated for your plug length and fluid densities.

Does the LERC4500 measure during rotation?

Yes. A digital camera images the receiving tube of each holder while the rotor turns, and the software converts the interface position into produced volume.

Is pore pressure available?

Pore pressure up to 1,450 psi is available with the overburden holders. Higher pore pressure is not part of the current scope.

Running capillary pressure on six plugs at once?

Send the rock type, plug size, fluid pair, target capillary pressure and temperature. We will match them to a rotor and configuration.

Request a quote

← Coretest Capillary Pressure Refrigerated Centrifuge LERC4500: description and specification

Also compare:

Trademark notice: Vinci Technologies and the model names referenced on this page are trademarks of their respective owners. Coretest is not affiliated with, authorized, endorsed or sponsored by these companies. Comparisons are based on each manufacturer's published brochures and product pages as of 28 September 2026 and are provided for engineering evaluation only. Specifications change; verify current data with each vendor before purchase.