Specification
Specification
| Core holders | 6; 1 in, 30 mm or 1.5 in plugs |
| Plug length | up to 4 in in atmospheric holders; up to 2.5 in in overburden holders (option) |
| Receiving tubes | 10 mL and 32 mL, calibrated |
| Speed | 50 to 4,500 rpm, regulation ±1 rpm, speed ramp |
| Rotor | outer radius 26 cm; up to 5,818 ×g |
| Temperature | 5 °C (at standstill) to 90 °C, refrigeration and heating, dual at-sample sensors |
| Readout | digital camera, 5488 × 3672 px; interface resolution better than 0.01 mL on a 10 mL tube |
| Tests | forced drainage and imbibition, air–liquid and oil–brine; production versus time for relative permeability |
| Overburden option | hydrostatic holders to 5,000 psi confining and 1,450 psi pore pressure, applied before the run |
| Protection | imbalance, rotor identification, overspeed, over-temperature, lid interlock |
| Power | 400 V, 3 phase + N |
| Size / weight | 973 × 813 × 1015 mm (H × W × D) / 450 kg |
Guaranteed values are stated in the project-specific technical proposal.